DTR3110-1064 Portable Raman Spectrometer
DTR3110-1064 portable Raman spectrometer is suitable for field operation. The outstanding reliability makes the detection result much more accurate. The excellent low stray light conditions that enable the spectrometer has a wide range of application, especially in public safety, food safety, and pharmaceutical engineering. The multi-function software facilitated the spectral analysis process in application and the remote experiment through internet access makes the test item much easier.
Features of the DTR3110-1064 Portable Raman Spectrometer
- High sensitivity 512 pixels InGaAs array; TE-cooled, down to -20℃,
- Ultra-low noise circuit;
- Powerful embedded software;
- Fluorescence background elimination;
- Peak value search and display;
- USB 2.0;
- User-friendly human and machine interface;
Application of Raman Spectrometer
- Biological science
- Material science
- Pharmaceutical engineering
- Forensic analysis
- Agriculture and food safety
- Gemstone
- Environmental science
Specifications of DTR3110-1064 Portable Raman Spectrometer
|
DTR3110-1064 System |
|
|
Interface |
USB2.0 |
|
Operating system |
Windows10 |
|
Integration time |
10ms-600000ms |
|
Power voltage |
DC 5V(+/-5%) |
|
Operating Temp |
-10〜40℃ |
|
Operating humidity |
< 95% |
|
Dimension(L*W*H) |
26x33x16.5cm |
|
Weight |
5.5Kg |
|
Reliability |
|
|
Spectral stability |
σ/μ < 0.5%(COT8hours) |
|
Temp stability |
Spectral shift< 1cm-1 (10-40℃) |
|
Variation of intensity (in5~40℃) |
<±5% |
|
Optical parameters |
|
|
一1 Spectral range (cm) |
200-2600 |
|
-1 resolution (cm) |
12-15 |
|
SNR |
>3000:1 |
|
Slit size |
50um, 25um optional |
|
Optical system |
f/4 C-T crossed optical path |
|
Max Quantum Efficiency |
>90% |
|
Focal length |
112mm for incidence and output |
|
Detector |
|
|
Item |
High sensitivity512pixels InGaAs array |
|
Cooled down to |
-20℃
|
|
Detecting range |
900-1700nm |
|
Effective pixels |
512 |
|
Dynamic range |
14,000: 1 |
|
Pixel size |
25×500 μm
|
|
Exciting Laser |
|
|
Central wavelength |
1064nm (+/-1nm) |
|
Semi-peak width |
0.1 nm |
|
Power output |
≥500mW |
|
Power stability |
σ/μ <±0.2% |
|
Raman probe |
|
|
Operating distance |
6mm |
|
Rayleigh scattering resistance |
OD>8 |
|
Numerical Aperture |
0.3 |
|
Aperture |
7mm |