DW-K5210-C-F Nano Spectrophotometer
The DW-K5210 Series is designed for micro-volume nucleic acid and protein measurement using only 0.3–2 μL of sample, with automatic optical path switching and results in less than 3 seconds. DW-K5210 provides ultra-micro measurement, DW-K5210C adds cuvette and OD600 detection, while DW-K5210F further adds fluorescence quantitation.
- Sample Volume: 0.3–2 μL
- Optical Paths: 1 / 0.4 / 0.05 mm, automatic switching
- Absorbance Range: 0–100 Abs
- dsDNA Range: 2–5000 ng/μL
- BSA Range: 0.06–149 mg/mL
- Detection Time: <3 s
Features of DW-K5210 Nano Spectrophotometer
Requires only a small sample volume for routine nucleic acid and protein testing.
Automatically switches between 1, 0.4 and 0.05 mm optical paths.
Measurement and zero adjustment start automatically when the detection arm is lowered.
DW-K5210C and DW-K5210F add cuvette measurement including OD600.
DW-K5210F adds fluorescence quantitation, standard curve and kinetics functions.
Measurement results can be stored in spreadsheet format and exported via USB.
Which Model Should I Choose?
- Choose DW-K5210 if:
you mainly need 0.3–2 μL micro-volume absorbance measurement for nucleic acid and protein samples.
- Choose DW-K5210C if:
you also require cuvette measurement and OD600 in addition to micro-volume testing.
- Choose DW-K5210F if:
you need micro-volume + cuvette + fluorescence measurement in one instrument.

Specifications
| Technical Parameters | |
| ★Optical path | Ultra-Micro Module:1mm、0.4mm、0.05mm (automatic optical path switching)Cuvette Module: 10mm, 5mm, 2mm, 1mm; |
| Volume requirements | Ultra-Micro Module: 0.3~2µLCuvette Module: Volume≥1000µL (for 10mm optical path standard cuvettes) or liquid height ≥8.5mm; |
| ★Detection Wavelength | Ultra-Micro Module: 230/260/280/340nmCuvette Module: 230/260/280/340/600 nm |
| Detection Repeatability | Ultra-Micro Module: ≤0.002A (1mm optical path)or ≤1%CV, whichever is larger;Cuvette Module: ≤0.002A (1mm optical path) or ≤1%CV, whichever is larger; |
| Absorbance Range | Ultra-Micro Module:0~100 Abs |
| Concentration Measurement Range | Ultra-Micro Module: 2~5000 ng/µL (dsDNA); 0.06~149 mg/mL (BSA); |
| Detection Time | Ultra-Micro Module: Less than 3 seconds |
| Ultra-Micro Module:Equipped with automatic measurement and automatic zero-adjustment functions; instant detection is achievable by lowering the detection arm;Cuvette Module:Equipped with automatic measurement and automatic zero-adjustment functions; instant detection is achievable by lowering the detection arm; ★Cuvette Detection Module Includes:Nucleic Acid Detection Module, Protein Detection Module, OD600 Module. | |
| Functional Parameters | |
| Ultra-Micro Module | dsDNA, ssDNA, RNA, User-defined, A280 |
| Cuvette Module | dsDNA, ssDNA, RNA, User-defined, A280, OD600 |
| Fluorescence Module | dsDNA, RNA, microRNA, RNA IQ, ssDNA, Protein Quantitation Kits, Ion Sphere, Fluorometer, Standard Curve, Kinetics |
| Hardware Parameters | |
| Display | 8-inch touch screen, multi-point capacitive touch |
| Operating System | LINUX |
| Supported Languages | 9 languages including Chinese, English, French, German, Arabic, Spanish, Russian, Japanese, and Korean; |
| Data Output | Supports USB data export and Ethernet connection |
| Data Output Ports | 2 USB-A ports, 1 Type-C port, 1 HDMI port, and 1 Ethernet port, enabling connection with various devices such as mice, keyboards, and desktop computers; |
| Data Storage | Measurement results are automatically saved in spreadsheet format and can be exported via USB. |
| Other Parameters | |
| Instrument Dimensions (L*W*H) | 230mm×320mm×190mm |
| Instrument Weight | 4 kg |
| Operating Conditions | Temperature10~45℃; Humidity: Below 85% |
| Operating Voltage | 24V 2.5A 60W |
Fluorescence
| Light source | Blue LED、 Red LED |
| Excitation filters | Blue excitation 430~495nm, Red excitation 590~640nm |
| Shutter filter | Green emission 510~580nm, Red emission 660~725nm |
| Detector | Photodiode (320~1100nm) |
| Sample throughput | 1 sample; |
| Sample volume | 1-20μL |
| Tube type | 0.5ml PCR tube |
| Tube Loading volume | ≥100μL |
| Preheating time | No preheating required |
| Measurement speed | ≤3s/tube |
| Linearity | R²>0.995 |
| Linear range | ≥ 5 orders of magnitude |
| Built-in Reagent Calculator: Calculates the required amounts of dye and buffer for working solutions; | |
| Includes Fluorometer Mode: Directly measures the fluorescence intensity of samples; | |
| Includes Ion Sphere Mode; | |
| ★Includes Fluorescence Standard Curve Module: Allows free selection of excitation and emission light sources, curve types, standard sample volume, and total reagent volume; | |
| ★Includes Fluorescence Kinetics Module: Supports customization of delay time, interval time, and number of detections for up to 5 stages. | |
| Model | DW-K5210 | DW-K5210C | DW-K5210F |
| Ultra-Micro Module | √ | √ | √ |
| Cuvette Module | / | √ | √ |
| Fluorescence Module | / | / | √ |
| Automatic detection | √ | √ | √ |
Applications of Nano Spectrophotometer

DNA & RNA Quantification
Measurement of dsDNA, ssDNA and RNA using micro-volume absorbance.

OD600 Measurement
DW-K5210C and DW-K5210F support OD600 cuvette measurement.

Protein Quantification
Supports A280 and compatible protein quantitation workflows.

Fluorescence Quantification
DW-K5210F supports fluorescence workflows for DNA, RNA, microRNA and compatible protein quantitation kits.
Ultra-Micro vs Cuvette Measurement
Ultra-Micro Module
Sample Volume: 0.3–2 μL
Optical Path: 1 / 0.4 / 0.05 mm
Wavelengths: 230 / 260 / 280 / 340 nm
Cuvette Module
Optical Path: 10 / 5 / 2 / 1 mm
Wavelengths: 230 / 260 / 280 / 340 / 600 nm
10 mm Standard Cuvette: ≥1000 μL or liquid height ≥8.5 mm
DW-K5210F Integrated Fluorescence Measurement
- dsDNA
- RNA
- microRNA
- RNA IQ
- ssDNA
- Protein Quantitation Kits
- Ion Sphere
- Fluorometer
- Standard Curve
- Kinetics