Coulometric Karl Fischer Titration
- Intelligent operation system provides features including user management, method management, sensor management, data management and GMP management etc.
- Support Auto Titration Method, Oven Titration Method, Oven Blank, and Coefficient correction Method.
- The solvent manager adopts anti-leakage liquid and anti-pour design of waste liquid bottle, supports to drain solution and feed solvent, ensure safe handling of Karl Fischer chemicals.
Includes of KFT-50C Karl Fischer Titrator
- Solvent manager
- Coulometric vessel
- Measuring electrode
- Generator electrode
Data Management
- Auto/manual drift adjust ensures accurate results.
- Selectable units including μg, mg, %, ppm, mg/L and μg/mL.
- User three-level authority management and password management.
- Support for GMP management, the GMP logs achieve the tracking and analysis of the the data.
- Store up to 2000 titration data sets (GLP-compliant).
- Support for RS-232 communication for print and the print format can be alternate.
- Support for USB communication for PC software.
- Support for U disk for method export and import.
- Support for U disk for data export with CSV or PDF format.
- Support for printer, barcode scanner, oven and balance (specific model).
- It needs GMP software to implement GMP management functions.
Specifications of KFT-50C Karl Fischer Titrator
| Model | KFT-50C | |
| Principle | Karl Fischer Coulometric Titration | |
| Water | Range | 3.0μg~200mg |
| Resolution | 0.1μg | |
| Unit | μg,mg,%,ppm,μg/mL,mg/L | |
| Repeatability | ≤0.3% | |
| mV Range | (0~2000)mV | |
| mV Resolution | 0.1mV | |
| Polarization Current Range | (1~200) μA | |
| Polarization Current Accuracy | ±3% | |
| Polarization Current Fluctuation | ±2.5%/30min | |
| Working Current Accuracy | ±0.5% | |
| Working Current Fluctuation | ±0.2%/10min | |
| Power Supply | (100~240)VAC,(47~63)Hz | |
| Dimensions | 240×370×270mm | |
| Net Weight | About 4kg | |