DU-8600RN Split Beam UV/VIS Spectrophotometer
- DU-8600RN UV/VIS spectrophotometers are advanced split beam ratio monitoring optical systems. They are suitable for biochemical, pharmaceutical, and clinical lab applications.
- Optional Accessories:
- PC scanning software
- Tungsten Lamp
- Deuterium lamp
- Micro cuvette cell holder (one-position)
- Manual 4-positions cell holder
- Automatic 4/8 positions cell holder
Features: 1. Split beam ensure low drift, low noise and Low stray light 2. High-speed MCU, high precision AD, large storage capacity 3. Large LCD display (320*240 Dots) 4. Data and Curve can be stored in real-time 5. Online software upgrade capability 6. Lamps can be turned on/off individually 7. Easy to change Pri-aligned lamps 8. Reinforced baseboard and bracket assures durability | Function: 1. Photometric 2. Quantitative(Standard Curve) 3. WL Scan(Spectrum Scan) 4. Time Scan(Kinetics) 5. DNA/Protein Test 6. Multi-WL Test 7. System Utility |
Specifications of DU-8600RN UV/vis Spectrophotometer
Model | DU-8600RN |
Wavelength Range | 190-1100nm |
Spectral Bandwidth | 2.0nm |
Optical System | Split Beam Ratio Monitoring, Blazed Holographic Grating (1200 lines/mm) |
Wavelength Accuracy | ±0.5nm |
Wavelength Repeatability | ≤0.2nm |
Wavelength Setting | Auto, Resolution 0.1nm |
Photometric Range | 0~200%T, -0.3~3A, 0~9999C |
Photometric Accuracy | ±0.002A (0~0.5A), ±0.003A (0.5~1A), ±0.3%T (0~100%T) |
Photometric Repeatability | ≤0.001A (0~0.5A), ≤0.002A (0.5~1A), ≤0.2%T (0~100%T) |
Stray Light | ≤0.05%T(220/360nm) |
Scan Speed | High, Medium, Low. Max.2000nm/minute |
Baseline Flatness | ±0.002A |
Stability | ±0.002A/h (500nm,0A) |
Noise | ≤0.2%T/3min (250/500nm,0%T); ≤0.5%T/3min (250/500nm,100%T) |
Sample Compartment | Accommodate 5-100mm Pathlength Cuvette |
Detector | Silicon Photodiode |
Lamps | Tungsten Lamp & Deuterium Lamp (Pre-aligned) |
Display | Graphic LCD (320*240 Dots) |
Keypad | 30-key Alphanumeric Membrance Keypad |
Output Port | USB Port & RS232 Port |
Printer | Mini Serial Printer; PC Printer |
PC Software | PC Scanning Software |
Power Requirements | AC 90-250V,50/60Hz |
Dimension | 545x468x245mm |
Weight | 18kg |
DU-8600RN Spectrophotometer Display

Customer Feedback of DU-8600RN Spectrophotometer

